Care
73
Disinfecting/cleaning/sterilising schedule for
Savina Intensive Care Ventilator
Applicable for use with non-infectious patients.
For infectious patients, all parts that conduct breathing gas
must be additionally sterilised after disinfecting and
cleaning.
The parts that conduct breathing gas listed here can be
steam-sterilised at 134
o
C. See "Sterilising" column.
This schedule contains guideline values only. The protocols
of the responsible hospital hygiene official remain
unaffected!
Part How often How
Reusable
components
Recommended
cleaning
intervals
Disinfecting and cleaning Sterilising
Autoclaving at
93
o
C 10 minutes
Wiping Bath
immersion
Steam
134
o
C
10 minutes
Savina
basic device
per patient no outside no no
Trolley
Hinged arm
Medical gas hose
per patient no outside no no
Ventilation hoses,
Y-piece,
Water traps and
Collecting jars
per patient/
weekly
yes no possible yes
Expiration valve per patient/
weekly
yes no possible yes
Flow sensor daily no*
* Special treatment, bath disinfect in 70 % ethanol, see page 70
outside possible* no
Temperature sensor daily no yes no yes