Preparation
96
Checklist
● To be carried out every time before using the device.
Device is complete and ready for operation.
The test lung is connected.
When the check has been completed successfully, Savina is
ready for use.
● Do not use device unless all the test steps have been
completed.
● Adapt all settings and alarm limits to current patient
situation.
What How What should happen
Operation of lamps/
LEDs and audible alarm
Switch device on: All lamps (except the lamps for power supply)
and LEDs light up, audible alarm sounds.
Operation of audible
alarm for mains power
failure
Hold down »Alarm Reset« for about
3 seconds:
Audible alarm sounds.
Apnoea alarm Set ventilation mode CPAP,
set PEEP 10 mbar,
steadily inflate and deflate the test lung to
simulate spontaneous breathing, then stop
simulation:
At the end of the apnoea alarm time T
Apnoea,
audible alarm sounds, the warning !!! Apnoea
appears on the screen.
Press »Alarm Reset«.
"PEEP high" alarm Set ventilation mode IPPV,
set Paw to 100 mbar,
keep flow sensor tight:
Audible alarm sounds, red lamp flashes,
the warning !!! PEEP high appears on the
screen.
Press »Alarm Reset«.
"Airway pressure high"
alarm
Set pressure limit Pmax to OFF, set Paw
below the current max. airway pressure:
Audible alarm sounds. The warning
!!! Airway pressure high appears on
the screen.
Set Paw to 100 mbar.
Press »Alarm Reset«.
If necessary, set pressure limit Pmax to ON.
"MV low" alarm Set MV higher than the current MV
measured value:
Audible alarm sounds, the red alarm lamp
flashes, the warning !!! MV low appears on
the screen.
Set MV lower than the current MV
measured value again.
Press »Alarm Reset«:
Tightness of breathing
system
Set high flow acceleration FlowAcc, press and
hold down the »Insp. hold« key and observe
pressure curve:
Plateau pressure remains constant.