Preparation
95
PEEP display:
PEEP 10 mbar
± 2 mbar tolerance
Minute volume display:
MV 7.8 L/min
± 1.0 L/min tolerance
O
2 concentration display:
FiO2 60 %
± 3 vol.% tolerance
Detach test lung from Y-piece
The following message is displayed:
»!!! Airway pressure low«.
Display after about 45 seconds:
MV 0 L/min
+ 0.5 L/min tolerance
Reconnect test lung to the Y-piece.
Checking mains power failure
Disconnect mains plug, or disconnect DC supply plug.
The device switches over to the internal battery and
continues ventilating.
Display:
!! Int. battery activated
Reconnect mains plug or DC supply plug.
Checking the O2 alarm
Disconnect O2 compressed gas hose connector:
Intermittent audible alarm and display:
!!! O
2 supply down
Savina ventilates with air.
Reconnect O2 compressed gas hose connector.
A warning !!! FiO2 low may appear briefly.
Press »Alarm Reset«.
After successful testing of readiness for operation, Savina is
ready to use.
Do not use device unless all the test steps have been
completed.