Technical data
122
Tachypnoea monitoring
Alarm if the total frequency is exceeded during spontaneous breathing.
Adjustment range 10 to 120/min
Apnoea alarm time
Alarm if no breathing activity is detected.
Adjustment range 15 to 60 s, adjustable in 1 second steps.
Insp. tidalvolume
Alarm if the delivered tidal volume V
T exceeds the alarm limit
Setting range 0.06 to 4 L
Operating data
Mains power connection 100 V to 240 V, 50/60 Hz
Current input
at 230 V max. 1.3 A
at 100 V max. 3.4 A
Power consumption in typical ventilation typically approx. 100 W
Machine fuses
Range 100 V to 240 V F 5 H 250 V IEC 127-2 (2x)
Protection class I, Type B
DC fuse for internal battery Plug-in fuse 1 SA/32 V
Supply from DC on-board network 12 to 36 V DC
Supply from external batteries 12 or 24 V
Input current (DC)
12 V battery typically 10 A, max. 20 A
24 V battery typically 5 A, max. 10 A
Bridging time if mains supply is down, with charged
external battery and typical ventilation
examples
12 V battery 36 Ah approx. 4 hours (e.g. with 1 lead-gel battery 12V/36 Ah)
24 V battery 17 Ah approx. 7 hours (e.g. with 2 lead-gel batteries 12V/17 Ah)
Bridging time if mains supply is down and no
external DC supply is available (new, charged
batteries)
approx. 60 minutes with typical ventilation
External battery
Sourcing information:
Standard rechargeable lead or lead-gel batteries
can be used, e.g.
Sonnenschein A212/36 A (1 or 2 batteries)
Panasonic LCL 12 V 17P (2 batteries)
LCL 12 V 33AP (1 or 2 batteries)
LCL 12 V 38P (1 or 2 batteries)