Technical data
121
Plateau time Tplat
Range 0 to 10 s
Resolution 0.1 s
Accuracy 0.1 s
Resistance R
Range 3 to 100 mbar/L/s
Resolution 1mbar/L/s
Accuracy ±2.5 mbar/L/s or 20 %, whichever is greater
Compliance C
Range 3 to 200 mL/mbar
Resolution 1 mL/mbar
Accuracy ±2 mL/mbar or 20 %, whichever is greater
Breathing gas temperature measurement
(NTC sensor)
Range 18 to 51
o
C
Resolution 1
o
C
Accuracy ±1
o
C
Curve displays
Airway pressure Paw (t) –5 to 100 mbar
Flow (t) –200 to 200 L/min
Monitoring
Expiratory minute volume MV
Upper alarm limit alarm when MV exceeds the upper alarm limit.
Setting range 41 to 2 L/min, in 0.1 L/min steps
Lower alarm limit alarm when MV falls below the lower alarm limit.
Setting range 0.5 to 40 L/min, in 0.1 L/min steps
Airway pressure Paw
Upper alarm limit alarm when the "Paw high" value is exceeded.
Setting range 10 to 100 mbar
Lower alarm limit alarm when the value "PEEP +5 mbar" (coupled with the PEEP set value)
is not exceeded for at least 0.1 s in two successive ventilation
strokes
Delay time TDisconnect for
"Airway pressure lower alarm limit" alarm
(only in »Mask/NIV« mode)
0 to 60 s
Insp. O2 concentration FiO2
Upper alarm limit alarm if FiO2 exceeds the upper alarm limit for at least 20 seconds.
Lower alarm limit alarm if FiO
2 falls below the lower alarm limit for at least 20 seconds.
Range both alarm limits are automatically allocated to the set value:
below 60 vol. % with ± 4 vol. %
above 60 vol. % with ± 6 vol. %