Flow measurement
(hot-wire sensor)
Inspiratory peak flow FlowPeak
Range 0 to 196 L/min
Resolution 1 L/min
Accuracy ±8 % of measured value
Minute volume MV
Spontaneous breathed minute volume MVspon
Range 0 to 99 L/min, BTPS*
* BTPS
Body Temperature, Pressure, Saturated.
Measured values with reference to the conditions of the patient’s lung,
body temperature 37
o
C, ambient pressure, steam-saturated gas.
Resolution 0.1 L/min
Accuracy ±8 % of measured value
T0...90 approx. 35 s
Leakage minute volume MVleak
based on the inspiratory minute volume
(only in »Mask/NIV« mode)
Range 0 to 100 %
Resolution 1 %
Accuracy ±18 % of measured value
Measured expiratory tidal volume VTe
Range 0 to 3999 mL, BTPS*
Resolution 1 mL
Accuracy ±8 % of measured value
Leakage-compensated inspiratory measured tidal
volume VTpat
(only in »Mask/NIV« mode)
Range 0 to 3999 mL, BTPS*
Resolution 1 mL
Accuracy ±18 % of measured value
Frequency Measurement
Spontaneous breathing frequency fspn
Total frequency ftot
Range 0 to 150 /min
Resolution 1 /min
Accuracy ±1 /min
T0...90 approx. 35 s
Inspiration/expiration ratio I:E
Range 0 : 150 to 150 : 1
Resolution 0.1
Accuracy ±6 % of measured value
Inspiration time Tinsp
Range 0 to 15 s
Resolution 0.1 s
Accuracy 0.1 s