Technical data
119
Performance data
Control principle Time-cycled, volume-constant, pressure-controlled
Intermittent PEEP frequency 2 cycles every 3 minutes
Medicament nebulisation
(with O
2 supply only)
For a maximum of 30 minutes, in the inspiratory flow phase,
2 bar, max. 10 L/min.,
Savina takes the nebulizer flow into account and keeps the
minute-volume constant
Oxygen enrichment for bronchial suction
(with O2 supply only)
disconnection detection automatic
reconnection detection automatic
oxygen enrichment max. 3 minutes with 100 vol.% O2
active suction phase max. 2 minutes
final oxygen enrichment 2 minutes with 100 vol.% O2
Supply system for spontaneous breathing and ASB Blower with quick-action pressure control valve
max. inspiratory flow 180 L/min, BTPS*
Equipment compliance
(with Fisher & Paykel MR 730 humidifier
and adult patient hose system) 2 mL/mbar
Inspiration resistance 2.3 mbar/60 L/min
Expiratory resistance 3.8 mbar/60 L/min
Equipment compliance
(with Fisher & Paykel MR 730 humidifier
and tubing system for paediatric use) 1 mL/mbar
Inspiration resistance 4.1 mbar/30 L/min
Expiration resistance 4.1 mbar/30 L/min
Additional functions
Inspiratory relief valve opens breathing system in case of failure
Safety valve opens the breathing system at 100 mbar.
Measured value displays
Airway pressure measurement
(resistive relative pressure sensor)
Max. airway pressure PPeak
Plateau pressure PPlat
Pos. end-exp. pressure PEEP
Mean airway pressure PMean
Range 0 to 99 mbar
Resolution 1 mbar
Accuracy ±2 mbar
Inspiratory O2 measurement:
(maintenance-free electrochemical sensor)
Inspiratory O2 concentration FiO2
Range 18 to 100 vol.%
Resolution 1 vol.%
Accuracy ±3 vol.%
* BTPS
Body Temperature, Pressure, Saturated.
Measured values with reference to the conditions of the patient’s lung,
body temperature 37
o
C, ambient pressure, steam-saturated gas.