Technical data
118
Technical data
Environmental conditions
In operation
Temperature 5 to 40°C
Atmospheric pressure 700 to 1060 hPa
Rel. humidity 5 to 95 %, no dew
In storage and transit
Temperature –20 to 70°C
Atmospheric pressure 600 to 1200 hPa
Rel. humidity 10 to 95 %, no dew
Settings
Ventilation modes IPPV/IPPVAssist,
SIMV, SIMV/ASB
BIPAP, BIPAP/ASB
CPAP/ASB
Ventilation frequency f 2 to 80 bpm
Inspiration time T
insp 0.2 to 10 s
Tidal volume VT 0.05 to 2.0 L, BTPS*
* BTPS
Body Temperature, Pressure, Saturated.
Measured values with reference to the conditions of the patient’s lung,
body temperature 37
o
C, ambient pressure, steam-saturated gas.
Accuracy ±10 % of set value or ±25 mL,
whichever is greater.
Inspiratory pressure Pinsp 0 to 99 mbar**
** 1 mbar = 100 Pa
O2 concentration
Accuracy 21 to 100 vol.%
±3 vol.%
The accuracy of the inspiratory O2 concentration is appreciably
reduced when operating Savina without O2 sensors.
Positive end-expiratory pressure PEEP
or interm. PEEP 0 to 35 mbar
Trigger sensitivity 1 to 15 L/min
Can be switched off in IPPV ventilation mode.
Pressure assist ∆ PASB with PEEP 0 to 35 mbar (relative to PEEP)
Flow Acceleration FlowAcc 5 to 200 mbar/s