Operation
27
CPAP, CPAP/ASB
Continuous Positive Airway Pressure
Assisted Spontaneous Breathing
For patients breathing spontaneously in an adequate manner.
Spontaneous breathing at a raised pressure level, to increase
the functional residual capacity FRC. Spontaneous breathing
can be assisted with additional pressure by ASB.
Set CPAP by means of the ventilation parameter keys:
—O2 concentration »O2«
Positive end-expiratory pressure »PEEP«
Additionally, for CPAP/ASB:
Pressure support »PASB above PEEP«
The following can be set on the screen:
FlowAcc for pressure support ASB on screen page
»Settings 1/1«
Sensitivity »Trigger« (for synchronising attempted
spontaneous breathing with ASB pressure support)
on screen page »Settings 1/1«.
CPAP, CPAP/ASB can be supplemented with the following
ventilation parameters and ventilation modes:
Trigger
By setting the trigger sensitivity, the assisting ventilator
support is synchronised with the patient’s own spontaneous
breathing attempts.
To set the trigger, see page 23.
Apnoea ventilation
To set apnoea ventilation, see page 28.
To set alarm limits, see page 30.