Operation
36
If an alarm limit has been switched off, an advisory message will
constantly be displayed, e.g.:
! MV low Alarm off
The apparatus does not display the following warning
messages in the »Mask/NIV« mode:
ASB > 4 s
Leakage
Setting T
Disconnect
A delay »TDisconnect« from 0 to 60 seconds can be set for the
»Airway pressure low« alarm in the »Mask/NIV« mode. In the
event of disconnection this delays the activation of the alarm by
the time set.
Press the »Alarms ∆∆« key until the screen page
»Alarms 2/2« appears.
Activate the »TDisconnect« line to set.
Set TDisconnect = turn the rotary knob,
confirm = press the rotary knob.
Selecting »Tube« mode
Switch apparatus to standby = hold down » Standby«
key for about 3 seconds.
Press »Alarm Reset« to switch off the audible alarm.
To select, activate »Patient Connection« line =
press rotary knob,
Select »Tube« = turn rotary knob,
Confirm = press rotary knob.
The advisory message appears briefly.
When switching over to »Tube« mode, Savina automatically
selects the default alarm limits for the alarms previously
switched off in »Mask/NIV« mode.
Check alarm limits after changing from »Mask/NIV«
mode to »Tube« mode.
129 37388130 37388