Operation
39
1 Hold down » Neb.« key, until the yellow lamp lights up.
On-screen advisory message:
! Nebulizer on
The nebulizer is in operation for 30 minutes; nebulization only
in the inspiratory flow phase.
If medicament nebulization needs to be stopped prematurely:
1 Press » Neb.« key again. The yellow lamp goes out,
the nebulizer is switched off.
The flow sensor will be automatically heat-cleaned and
calibrated by Savina.
On-screen information:
Flow calibration
Remove any left-over medicament from the medicament
nebulizer.
Follow the instructions for use of the medicament nebulizer.
Take into account the effect of aerosols on sensors and
filters.
The measuring function of the flow sensor may be impaired.
The flow resistance of filters is liable to increase and may impair
ventilation.
During medicament nebulization do not use a heat and
moisture exchanger (HME). Risk of increased
breathing resistance.
Do not place microbial filters on the nebulizer outlet
during nebulization!
Risk of increased breathing resistance!