Operation
41
After suction disconnection
During the period of disconnection Savina supplies a minimal
flow of 100 Vol.% O
2 in order to automatically detect the end of
the disconnection phase. The time available for suction is
continuously displayed in the information line on screen
(example):
Execute suction and reconnect 120 s
If suction is ended and the system is reconnected within the
displayed time, Savina starts the final O2 enrichment.
After reconnection
After reconnection has taken place, Savina ventilates once
more in the set ventilation mode, however, for final O2
enrichment 120 seconds more with 100 Vol.% O2.
Message in the information line on screen:
Final O2 enrichment 100% 120 s
The remaining time is continuously displayed on screen.
In case the O2 enrichment has to be interrupted:
1 Press the »O2 Y suction« key again
Automatic discontinuation of oxygen enrichment
If after 120 seconds no reconnection follows, Savina stops the
O2 enrichment program.
All alarms are immediately reactivated. Savina immediately
continuous ventilating in the set ventilation mode.