Scanning
48 EchoNous Bladder & Vein User Guide
FIGURE 34. Distance Measurement
Enabling depths of 4 and 5 CM
On EchoNous Vein, there are three ways of enabling depths of 4 and 5 CM, if desired.
To enable the setting for all scans:
1. From the home screen, tap the icon in the upper right corner.
2. Tap Settings.
3. Tap Vascular Access.
4. Select Enable 4 and 5 CM depth.
To enable the setting before a scan and for only the one scan:
1. Tap the icon in the upper right corner.
2. Select Enable 4 and 5 CM depth (see FIGURE 35. Enable 4 and 5 CM depth).
FIGURE 35. Enable 4 and 5 CM depth
3. Tap the Scan icon to start the scan.
To enable the setting during the scan and for only the one scan:
1. Tap the Stop icon to freeze the image.
2. Tap the icon in the upper right corner.
3. Select Enable 4 and 5 CM depth.
4. Tap the Scan icon to continue the scan.
Vascular access measurement
Tap the Measurement icon to insert calipers, arrows, or volume measurements on images: