EchoNous Bladder & Vein User Guide 71
System connectivity
EchoNous AI Station troubleshooting
If the System does not wake up after following the wake up instructions, disconnect and then reconnect the
USB cable connecting the HUB to the tablet display.
If this is unsuccessful, connect the power cord to an AC source.
If the System is still not responding, restart the tablet display by holding down the power button and
selecting Restart.
If the System remains unresponsive or a HUB fault error is displayed, please contact EchoNous
support.
Cleaning
EchoNous recommends the use of a soft cloth, lightly dampened with warm water or a mild soap solution for
cleaning the AI Station. Wipe surfaces with the moist cloth, and towel dry with a soft clean cloth.
Disinfection
Where bodily fluid or other contamination is expected, the EchoNous stands can be disinfected using the spray
and wipe method with a variety of disinfection products. Follow the manufacturers recommended method.
The following disinfecting means can be used with the MODO Stand:
Sani-Cloth
®
Active by PDI of Flint, UK
Tuffie 5 Wipes by VernaCare of Lancashire, UK
Protex Disinfectant Wipes by Parker Laboratories of New Jersey
Any Isopropyl alcohol impregnated wipes with <= 70% w/w alcohol
The following disinfecting means can be used with the AI Station and AI Station 2:
Clorox Hydrogen Peroxide wipes
Clorox Bleach wipes
PDI Sani-Cloth Plus
Refer to EchoNous
website for additional disinfection options.
1. First, clean the mobile stand as above.
2. Wipe or spray the mobile stand with disinfectant, and leave for the disinfectant manufacturers recommended
contact time.
3. Remove any residue with a soft cloth moistened with water.