Specifications
74 EchoNous Bladder & Vein User Guide
Environmental operating and storage conditions
The EchoNous System is intended to be used and stored in normal ambient conditions inside a medical facility.
Mode of Operation
The System enforces scanning limits when the display battery is at or below 5%.
Power supplies (chargers)
For the Samsung tablet display, the power supply is EchoNous Type C power supply (P005179).
Input: 100-240V AC supply 50-60Hz
Output: 5.0V ±0.25V at 2.0A maximum current
Internal battery
For the Samsung S3 tablet display, the battery is a 6000 mAh Li-ion polymer type.
For the Samsung S6 tablet display, the battery is a 7040 mAh Li-ion type.
TABLE 4. Operating, charging, transport, and storage condition ranges
Operating Transport and Storage
Temperature (°C) 5 (41°F) – 40°C (104°F) -20 (-4°F) – 50°C (122°F)
Relative humidity (non-condensing) 15–90% 10–93%
Pressure 70–101.5 kPa (10.3-14.7 psi) 50–101.5 kPa (7.3-14.7 psi)
TABLE 5. Charging condition ranges
Charging
Temperature (°C) 5 (41°F) – 40°C (104°F)
Relative humidity (non-condensing) 20–80%
Pressure 70–101.5 kPa (10.3-14.7 psi)
After storing at extreme temperatures, check the transducer surface
temperature before applying to a patient. A cold or hot surface may burn
a patient.
Only operate, charge, and store the System within the approved
environmental parameters.
40 °C (104°F)
5°C (41°F)
50 °C (122°F)
-20°C (-4°F)
40 °C (104°F)
5°C (41°F)