EchoNous Bladder & Vein User Guide 81
Electromagnetic compatibility (EMC)
Electromagnetic compatibility (EMC)
The System is intended for use in the electromagnetic environment specified below. The user of the System
should assure that it is used in such an environment.
1
The System has Class A compliance in meaning it is suitable for use in all establishments other than domestic and those
directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. If
the System is found to cause or respond to interference follow the guidelines in the warning section above.
The System complies with the Electromagnetic Compatibility
requirements of AS/NZ CISPR 11:2015 and EN IEC 60601-1-2:2014.
However, electronic and mobile communications equipment may
transmit electromagnetic energy through air and there is no guarantee
that interference will not occur in a particular installation or environment.
Interference may result in artifacts, distortion, or degradation of the
ultrasound image. If the System is found to cause or respond to
interference, try re-orienting the System or the affected device, or
increasing the separation distance between the devices. Contact
EchoNous customer support or your EchoNous distributor for further
information.
The System contains sensitive components and circuits. Failure to
observe proper static control procedures may result in damage to the
System. Any faults should be reported to
EchoNous or your EchoNous
distributor for repair.
TABLE 1. Guidance and manufacturer's declaration: electromagnetic emissions
Emissions test Compliance Electromagnetic environment: guidance
RF emissions
CISPR 11
Group 1 The System uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The System is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies