M3 Vital Signs Monitor User Manual NIBP Monitoring (Optional)
- 68 -
Message
Cause
Alarm Level
What to do
OVER
PRESSURE
Pressure has exceeded the
specified upper safety limit.
Low
Measure again, if failure
persists, stop using
measuring function of NIBP
module and notify
biomedical engineer or
manufacturer’s service staff.
NIBP
SECONDAR
Y OVER
PRESSURE
Secondary over pressure
protection
High
Notify biomedical engineer
or manufacturer’s service
staff.
CUFF TYPE
ERR
The cuff type used isn’t
consistent with the patient
type.
Low
Confirm the patient type
and change the cuff.
NIBP TIME
OUT
Measuring time has
exceeded the specified time.
Low
Measure again or use other
measuring method.
Prompt message: (display in the prompt area below NIBP value):
Message
Cause
Manual measuring…
During manual measuring mode.
Continual Measuring
In continuous measuring mode.
Auto measuring…
During automatic measuring mode.
Measurement over
Measurement over
Calibrating…
During calibrating
Calibration over
Calibration over
Leak testing…
During leakage test
Leak test over
Leakage test is over
Resetting...
NIBP module in resetting