M3 Vital Signs Monitor User Manual TEMP Monitoring (Optional)
- 84 -
Message
Cause
Alarm Level
What to do
Temp Error P03
Monitor mode heater
thermistor unstable or
out of range.
Low
Temp Error P04
Predict Mode patient
thermistor unstable or
out of range.
Low
Temp Error P05
Predict Mode heater
thermistor unstable or
out of range.
Low
Temp exceed
limit
The TEMP value is
out of the range of
+30 C ~ +43 C.
Medium
Put the probe into the probe
well; take it out and measure
again.
No TEMP
SENSOR
Probe configuration
(or no probe
connected) error.
Low
Well connect the probe and the
monitor, and measure again.
Temp Error P06
Unable to pre-heat
probe tip.
Low
Stop using measuring function
of TEMP module; notify
biomedical engineer or
Manufacturer’s service staff.
NOTE: Measure readings
displayed on the screen are
unreliable when the monitor
indicates Temp Error P06.
Measure Site
Error
The probe in use is
not consistent with
the measure position
set on the monitor.
Medium
Correctly set the measure
position on the monitor.
Prompts:
Message
Cause
Warm-up over
The monitor prompts it after the probe is taken out of the
probe well and warm-up is over.
Measure over
Prediction measurement is completed.
Quick Predict Over
Quick prediction measurement is completed.
MONITORING
The mode is switching to Monitor Mode.