SCU-800 Control Unit for Turbomolecular Pump
MT-49E-101-G
Page 82
5
SERIAL COMMUNICATION
PROTOCOL
5.4.2 ReadMeas
Function: Reads the measured rotational speed.
Transmission frame:
PC->SIM
Stx
0
0
1
?
D
Etx
LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC
Stx
0
0
1
Sp
D
Parameter 1 to 2
Etx
LRC
Parameter
Item
Data format
Remarks
1
[System reservation]
56-bits hexadecimal
coded ASCII
2
Measured rotational speed
(Unit: Hz)
16-bits hexadecimal
coded ASCII
Example:
Measured rotational speed: 02DC
hex
= 732 Hz = 43,920 rpm
Parameter
1
2
ASCII
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
0
2
D
C
HEX
30
32
44
43
*1
System reservation
5.4.3 Command
Function: Sends the pump operation commands START, STOP and RESET. These commands
function in the same way as when each switch of the SCU-800 is pressed. They are valid when
MANUAL/REMOTE changeover switch on the SCU-800 is "ON". Refer to Section 3.11, "Parameter
set mode".
Transmission frame:
PC->SIM
Stx
0
0
1
Sp
E
Parameter 1
Etx
LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC
Stx
0
0
1
#
Etx
LRC