How to use
42
Selected gas: none, oxygen, nitrogen
Balance purging time in seconds
If checked lids are closed during this stage
*)
Will be shown depending on the selected stage type.
Stop by maximum time
This stage is finished when specified Time, counted after reaching setpoint, is expired.
Stage result is the last weight before stage ends.
Fig.
: Stop by maximum time
Stop by time
This stage is finished when specified Time, counted after reaching setpoint, is expired and
additionally one more weight for all crucibles is taken.
The software will interpolate sample weight at specified time to obtain stage result.
Fig.
:Stop by time