94
Reference Manual
00809-0100-4840, Rev CB
Appendix B: Product Certifications
February 2015
Product Certifications
B.4.3 International certifications
NEPSI China intrinsically safe approval (on the 3105 only)
I3 Certificate: GYJ081008X
NEPSI Intrinsic Safety
Ex ia IIC T6 (T
a
= –40 to 55 °C)
Ex ia IIC T4 (T
a
= –40 to 60 °C)
Ui = 30 V, li = 120 mA, Pi = 0.82 W, Li = 108 H, Ci = 0 F
(see “Instructions specific to hazardous area installations (I3)” on page 96)
IECEx intrinsically safe approval (on the 3105 only)
I7 Certificate: IECEx SIR 06.0068X
IECEx Intrinsic Safety
Zone 0, Ex ia IIC T6 Ga (T
a
= –40 to 55 °C)
Zone 0, Ex ia IIC T4 Ga (T
a
= –40 to 60 °C)
Ui = 30 V, li = 120 mA, Pi = 0.82 W, Li = 108 H, Ci = 0 F
(see “Instructions specific to hazardous area installations (I1 and I7)” on page 95)