Modbus Commands
9.4 List of Parameters and Registers - Sorted by Daniel Registers
Address Tag Name
Data
Type
Client
Access
Description
Daniel Modicon
8043
5085…
5086
Status.NamurStates.MaintRequAl-
armComp5
DWord RO
Namur MaintRequ alarms that are component5
related
8044
5087…
5088
Status.NamurStates.FctCheckA-
larmComp5
DWord RO
Namur FctCheck alarms that are component1
related
8181
5361…
5362
Status.IO.ValveState DWord RO Valve state biteld (b0=V1, b1=V2, ....b19=V20)
9001
6001…
6002
PV1 Float RO Primary Variable 1
9002
6003…
6004
PV2 Float RO Primary Variable 2
9003
6005…
6006
PV3 Float RO Primary Variable 3
9004
6007…
6008
PV4 Float RO Primary Variable 4
9005
6009…
6010
PV5 Float RO Primary Variable 5
9006
6011…
6012
Service.Measurement.Simulation.
PVARawValue1
Float RO value for rawPVA of comp1
9006
6011…
6012
Status.Diagnostics.Raw-
QuotConce1
Float RO raw ADC quotient of component1
9007
6013…
6014
Service.Measurement.Simulation.
PVARawValue2
Float RO value for rawPVA of comp2
9007
6013…
6014
Status.Diagnostics.Raw-
QuotConce2
Float RO raw ADC quotient of component2
9008
6015…
6016
Service.Measurement.Simulation.
PVARawValue3
Float RO value for rawPVA of comp3
9008
6015…
6016
Status.Diagnostics.Raw-
QuotConce3
Float RO raw ADC quotient of component3
9009
6017…
6018
Service.Measurement.Simulation.
PVARawValue4
Float RO value for rawPVA of comp4
9009
6017…
6018
Status.Diagnostics.Raw
-
QuotConce4
Float RO raw ADC quotient of component4
9010
6019…
6020
Service.Measurement.Simulation.
PVARawValue5
Float RO value for rawPVA of comp5
9010
6019…
6020
Status.Diagnostics.Raw-
QuotConce5
Float RO raw ADC quotient of component4
9011
6021…
6022
Status.Diagnostics.RawMeas-
Conce1
Float RO raw ADC of measure-side component1
9012
6023…
6024
Status.Diagnostics.RawRefConce1 Float RO raw ADC of reference side component1
9013
6025…
6026
Status.Diagnostics.RawMeas-
Conce2
Float RO raw ADC of measure-side component2
9014
6027…
6028
Status.Diagnostics.RawRefConce2 Float RO raw ADC of reference side component2
9015
6029…
6030
Status.Diagnostics.RawMeas-
Conce3
Float RO raw ADC of measure-side component3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM XE and is the answer not in the manual?

Emerson X-STREAM XE Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM XE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals