7.2 General Maintenance Information
7.2 General Maintenance Information
Visual Inspections
Component Interval
Tubing, exible Leakage, embrittlement, contamination
Once a year
Tubing, stainless steel (SS) Corrosion, contamination
Pressure sensor, pressure switch,
Flowmeter
Corrosion, leakage
Pump Fixed screws, swing free to move
Valve block Corrosion, leakage
Flame arrestors Corrosion, damages, rmly seated
Field housings (IP 66 / NEMA 4X) Corrosion, damages on enclosure and
gaskets
Field housings stopping plugs Firmly seated
Field housings cable glands Firmly seated
Intervals given in the following tables are
based on standard operating conditions
(ambient temperatures +10 ... +40 °C /
+50 ... +104 °F; temperature changes < 10 K
/hr).
Try cleaning contaminated components.
Replace components showing corrosion, or
not passing inspections or tests!
Maintenance intervals must be
shortened for differing opera-
ting conditions, and if aggres-
sive gases are supplied.
Take care of special mainte-
nance instructions in separate
manuals for accessories or
safety equipment, e.g. ame ar-
restors, infallible containments,
etc.
If applicable, consider the ma-
nual addendums for instru-
ments for hazardous areas!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Emerson X-STREAM XE and is the answer not in the manual?

Emerson X-STREAM XE Specifications

General IconGeneral
BrandEmerson
ModelX-STREAM XE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals