EasyManua.ls Logo

Emerson X-STREAM XE - Cleaning the Instrument´s Outside

Emerson X-STREAM XE
536 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Emerson Process Management GmbH & Co. OHG7-62
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
7.7 Cleaning the Instrument´s Outside
7.7 Cleaning the Instrument´s Outside
Use a liquid general purpose detergent and
a lint-free cloth for cleaning the analyzer´s
outside.
Seal the open analyzer´s gas
ttings utilizing PVC caps to
avoid contamination of inner
gas path.
Moisten the lint-free cloth with a mixture of
3 parts of water and 1 part of the general
purpose detergent.
Procedure
Disconnect the instrument from power!
If disconnecting from gas lines is required,
take care of the following:
HAZARD FROM UNHEALTHY SUBSTANCES
Take care to follow the safety instructions and instructions for use given
by the manufacturer of the chosen general purpose detergent!
Do NOT drench the cloth, just
moisten it to prevent liquid en-
tering the housing!
Clean the analyzer housing outside with
the moistened cloth.
If need be dry the housing after cleaning.
EXPLOSIVE, FLAMMABLE AND HARMFUL GASES HAZARD
Before opening gas paths they must be purged with ambient air or neutral
gas (N
2
) to avoid hazards caused by toxic, ammable, explosive or harmful
to health gas components!

Table of Contents