Proline Promag 55 PROFIBUS DP/PA Troubleshooting
Endress+Hauser 127
061 S: HW F-CHIP
$: # 061
0x0F BAD
BAD
Device Failure
Device Failure
Constant F-CHIP defect/not
plugged
Cause:
F-Chip is not plugged into the I/O board or is
missing.
F-Chip is defective
Remedy:
1. Check whether the F-Chip is correctly plugged
into the I/O board.
2. Replace the F-CHIP.
3. Plug the F-CHIP into the I/O board.
No. # 1xx Software error
101 S: GAIN ERROR AMP.
$: # 101
0x0F BAD
BAD
Device Failure
Device Failure
Constant Gain Error
Amplifier
Cause:
Gain deviation compared to reference gain is
greater than 2%.
Remedy:
Replace the amplifier board ä 135.
121 S: A/C COMPATIB.
!: # 121
0x0F BAD
BAD
Device Failure
Device Failure
Constant Amplifier and I/O
board only partially
compatible
Cause:
Due to different software versions, I/O board and
amplifier board are only partially compatible
(possibly restricted functionality).
!
Note!
The indication on the display as notice message
appears only for 30 seconds (with listing in
"Previous system condition" function).
This condition can occur if only one electronics
board has been exchanged; the extended
software functionality is not available. The
previously existing software functionality is still
working and the measurement possible.
Remedy:
Module with lower software version must either be
updated by FieldCare using the required
(recommended) software version, or the module
must be replaced ä 135.
No. # 2xx Error in DAT / no communication
205 S: LOAD T-DAT
!: # 205
0x0F BAD
BAD
Device Failure
Device Failure
Constant Save to T-DAT
failed
Cause:
Data backup (downloading) to T–DAT failed, or
error when accessing (uploading) the calibration
values stored in the T–DAT.
Remedy:
1. Check whether the T-DAT is correctly plugged
into the amplifier board.
2. Replace the T-DAT if it is defective ä 135.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics. Check the:
Spare part set number
Hardware revision code
3. Replace measuring electronics boards if
necessary ä 135.
206 S: SAVE T-DAT
!: # 206
0x0F BAD
BAD
Device Failure
Device Failure
Constant Restore from T-
DAT failed
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
(Spare parts ä 135 ff.)