Troubleshooting Proline Promag 55 PROFIBUS DP/PA
128 Endress+Hauser
261 S: COMMUNICAT. I/O
$: # 261
0x18
0x19
0x1A
BAD
BAD
No
Communica-
tion
O.K.
Low
High
Communication
failure
Cause:
Communication error. No data reception between
amplifier and I/O board or faulty internal data
transfer.
Remedy:
Check whether the electronics boards are correctly
inserted in their holders ä 137.
No. # 3xx System limits exceeded
321 S: TOL. COIL CURR.
$: # 321
0x0F BAD
BAD
Device Failure
Device Failure
Constant Coil current out of
tolerance
Cause:
The coil current of the sensor is out of tolerance.
Remedy:
#
Warning!
Switch off power supply before manipulating the
coil current cable, coil current cable connector or
measuring electronics boards!
Remote version:
1. Check wiring of terminals 41/42 ä 40.
2. Check coil current cable connector.
Compact and remote version:
Replace measuring electronics boards if necessary
339
342
S: STACK CUR OUT n
$: # 339 to 342
0x0F BAD
BAD
Device Failure
Device Failure
Constant Temp. Buf. not
cleared
Cause:
The temporarily buffered flow portions (measuring
mode for pulsating flow) could not be cleared or
output within 60 seconds.
Remedy:
1. Change the upper or lower limit setting, as
applicable.
2. Increase or reduce flow, as applicable.
Recommendation:
Configure the fault response of the output to
ACTUAL VALUE, so that the temporary buffer
can be cleared.
Clear the temporary buffer by the measures
described under Item 1.
343
346
S: STACK FREQ. OUT n
$: # 343 to 346
0x0F BAD
BAD
Device Failure
Device Failure
Constant Temp. Buf. not
cleared
347
350
S: STACK PULSE OUT n
!: # 347…350
0x0F BAD
BAD
Device Failure
Device Failure
Constant Temp. Buf. not
cleared
Cause:
The temporarily buffered flow portions (measuring
mode for pulsating flow) could not be cleared or
output within 60 seconds.
Remedy:
1. Increase the setting for pulse weighting.
2. Increase the max. pulse frequency, if the
totalizer can handle a higher number of pulses.
3. Increase or reduce flow, as applicable.
Recommendation:
Configure the fault response of the output to
ACTUAL VALUE, so that the temporary buffer
can be cleared.
Clear the temporary buffer by the measures
described under Item 1.
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
(Spare parts ä 135 ff.)
Quality code (HEX)
Measured value status
Quality status
Quality substatus
Limits