Proline Promag 55 PROFIBUS DP/PA Troubleshooting
Endress+Hauser 129
351
…
354
S: CURRENT SPAN n
!: # 351…354
0x54
0x55
0x56
UNCERTAIN
UNCERTAIN
Engineering
Unit Range
Violation
Engineering
Unit Range
Violation
O.K.
Low
High
Flow is Out of
Range
Cause:
Current output: The actual value for the flow lies
outside the set limits.
Remedy:
1. Change the upper or lower limit setting, as
applicable.
2. Increase or reduce flow, as applicable.
355
…
358
S: FREQ. RANGE n
!: # 355…358
0x54
0x55
0x56
UNCERTAIN
UNCERTAIN
Engineering
Unit Range
Violation
Engineering
Unit Range
Violation
O.K.
Low
High
Flow is Out of
Range
Cause:
Frequency output: The actual value for the flow lies
outside the set limits.
Remedy:
1. Change the upper or lower limit setting, as
applicable.
2. Increase or reduce flow, as applicable.
359
…
362
S: PULSE RANGE
!: # 359…362
0x54
0x55
0x56
UNCERTAIN
UNCERTAIN
Engineering
Unit Range
Violation
Engineering
Unit Range
Violation
O.K.
Low
High
Flow is Out of
Range
Cause:
Pulse output: The pulse output frequency is out of
range.
Remedy:
1. Increase the setting for pulse weighting.
2. When selecting the pulse width, choose a
value that can still be processed by a connected
counter (e.g. mechanical counter, PLC, etc.).
Determine the pulse width:
– Version 1: Enter the minimum duration that
a pulse must be present at the connected
counter to ensure its registration.
– Variant 2: Enter the maximum (pulse)
frequency as the half "reciprocal value" that
a pulse must be present at the connected
counter to ensure its registration.
Example:
The maximum input frequency of the
connected counter is 10 Hz. The pulse width
to be entered is:
A0004437
3. Reduce flow
No. # 5xx Application error
501 S: SW.-UPDATE ACT.
!: # 501
0x48
0x49
0x4A
UNCERTAIN
UNCERTAIN
Substitute set
(substitute set
of failsafe
status)
O.K.
Low
High
New amplifier
software loaded
Cause:
New amplifier or communication software version
is loaded.
Currently no other functions are possible.
Remedy:
Wait until the procedure is finished. The device
will restart automatically.
No. Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
(Spare parts ä 135 ff.)
Quality code (HEX)
Measured value status
Quality status
Quality substatus
Limits