Commissioning Proline Promass F 300 EtherNet/IP
146
Parameter Prerequisite Description Selection Factory setting
Totalizer operation mode One of the following options is
selected in the Assign process
variable parameter
(→ 145) of the Totalizer
1 to n submenu:
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow
*
• Carrier mass flow
*
Select totalizer calculation
mode.
• Net flow total
• Forward flow total
• Reverse flow total
Net flow total
Failure mode One of the following options is
selected in the Assign process
variable parameter
(→ 145) of the Totalizer
1 to n submenu:
• Volume flow
• Mass flow
• Corrected volume flow
• Target mass flow
*
• Carrier mass flow
*
Define totalizer behavior in
alarm condition.
• Stop
• Actual value
• Last valid value
Stop
* Visibility depends on order options or device settings