Proline Promass F 300 EtherNet/IP Diagnostics and troubleshooting
207
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
834 Process temperature too high Reduce process temperature 0x80000C5
Status signal S
Diagnostic behavior [from the
factory]
1)
Warning
1) Diagnostic behavior can be changed.
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
835 Process temperature too low Increase process temperature 0x80000C6
Status signal S
Diagnostic behavior [from the
factory]
1)
Warning
1) Diagnostic behavior can be changed.
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
842 Process limit Low flow cut off active!
1. Check low flow cut off
configuration
0x8000091
Status signal S
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
862 Partly filled pipe 1. Check for gas in process
2. Adjust detection limits
0x8000092
Status signal S
Diagnostic behavior Warning
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
882 Input signal 1. Check input configuration
2. Check external device or process
conditions
0x1000031
0x1000257
Status signal F
Diagnostic behavior Alarm