Diagnostics and troubleshooting Proline Promass F 300 EtherNet/IP
208
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
910 Tubes not oscillating 1. Check electronic
2. Inspect sensor
0x1000050
Status signal F
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
912 Medium inhomogeneous 1. Check process cond.
2. Increase system pressure
0x80000C4
0x80000DF
0x8000115
0x8000162
Status signal S
Diagnostic behavior [from the
factory]
1)
Warning
1) Diagnostic behavior can be changed.
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
913 Medium unsuitable 1. Check process conditions
2. Check electronic modules or
sensor
0x80000CD
Status signal S
Diagnostic behavior [from the
factory]
1)
Alarm
1) Diagnostic behavior can be changed.
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
941 API temperature out of specification 1. Check process temperature with
selected API commodity group
2. Check API related parameters
0x8000380
Status signal S
Diagnostic behavior Alarm
Diagnostic information Remedy instructions Coding of diagnostic
information (hex)
No. Short text
942 API density out of specification 1. Check process density with
selected API commodity group
2. Check API related parameters
0x800033B
Status signal S
Diagnostic behavior Alarm