5 System settings and "Service" menu
18 / 84
80116-938_V24392
2023-03
Argon
plasma
module APC
V 1.0.x Automatic irrigation:
Once it has been plugged into the APC socket, the instrument is auto-
matically irrigated with gas / an instrument that has already been
plugged into the APC socket is automatically irrigated with gas when
the unit starts up.
ON or OFF
V 1.0.x Irrigation time:
Selection of the time period during which the instrument will be auto-
matically irrigated with gas.
0 to 10 seconds
V 1.0.x Gas supply:
Gas bottle or central gas supply
V 1.0.x Pressure with full gas bottle:
This setting determines the maximum bottle pressure of the argon gas
bottle in use.
The correct output of the bottle fill level on the display of the electro-
surgical unit depends on the maximum bottle pressure set for the
respective argon gas bottle that is in use.
100 to 240 bar
IES smoke
evacuator
Only for V 1.0.x. Basic suction:
Setting of the smoke evacuator during breaks between working. The
system works with reduced power during basic suction.
OFF or 1 to 30 %
Only for V 1.0.x. Run-on time:
Setting of the smoke evacuator for the period of time following can-
celation of activation.
OFF, 1 to 10 seconds
Protected
settings
V 1.0.x Password-protected access to level 2 for changing system settings.