19 / 84
5 • System settings and "Service" menu
80116-938_V24392
2023-03
Overview of system settings 2:
Protected settings
Only accessible for service personnel. Password-protected.
Setting Available
from
Description
Maximum activa-
tion time
V 1.0.x This setting determines the period of time after which activation auto-
matically terminates.
OFF or 3 to 99
Save changed pro-
grams
V 1.0.x This setting determines whether changes made in a program can be
saved in such a way that they are retained when the electrosurgical
unit is switched off.
allowed
blocked
Assignment of two
pedal footswitch
V 1.0.x This setting determines whether the pedals of a two pedal footswitch
may be used to activate different instruments (CUT for Instrument 1,
COAG for Instrument 2).
allowed
blocked
Use of AUTO START V 1.0.x Function whereby the electrosurgical unit is automatically activated
(immediately or following a delay) when an instrument has contact
with tissue:
allowed
blocked
Delay AUTO START
bipolar
V 1.0.x Input of start delay for the AUTO START function in bipolar mode.
0 to 9 seconds
Delay AUTO START
thermoSEAL
V 1.0.x Input of start delay for the AUTO START function in thermoSEAL mode.
0 to 9 seconds
Preferred return
electrode type
V 1.0.x This setting determines the return electrode used.
non-split
split