17 / 84
5 • System settings and "Service" menu
80116-938_V24392
2023-03
CHAPTER 5
System settings and "Service" menu
General information
System settings This unit has two levels for changing system settings. The first level can be
accessed by users and service personnel. The second level can only be ac
-
cessed by service personnel.
"Service" menu The "Service" menu offers test programs, various lists for fault analysis
and other unit information relevant for service. It is only intended for ser
-
vice personnel.
Overview of system settings 1:
Further settings
Accessible to users and service personnel.
Setting Available from Description
Volume V 1.0.x Stepless adjustment of the unit volume:
Overall: All unit sounds together
Individually: Activation sounds, message sounds or key sounds
IMPORTANT! For safety reasons, message sounds are always output
in clearly audible fashion by the unit (regardless of the setting actually
selected).
Brightness V 1.0.x Stepless adjustment of the unit brightness:
Overall: All unit displays together
Individually: Screen or socket frame
Date V 1.0.x Self-explanatory.
Time V 1.0.x Self-explanatory.
Language V 1.0.x Self-explanatory.
Start screen V 1.0.x Selection of the start screen:
Program group list (e.g. gynecology, general surgery, etc.),
Program list (overview of last programs used)
FocusView (last program used)