EIP/EIP-A error
Display on the unit: EIP or EIP-A, detecting module: Erbe irrigation pump EIP 2 (from V 1.1.x)
Error code Description Additional information Action
EIP Not available.
EIP-A 128 CRC error. Restart the unit. If the
error occurs again, notify
Erbe Service.
EIP-A 130 Pump cover opened when
switched ON.
Close pump cover before
activation and/or do not
open during activation.
Restart the unit. If the
error occurs again, notify
Erbe Service.
EIP-A 132 Timeout error. End the activation. It is
then possible to reactivate
the EIP 2. If the error
occurs again, check the
accessories.
EIP-A 136 UeWS error. Restart the unit. If the
error occurs again, notify
Erbe Service.
EIP-A 144 No electricity and/or con-
sumer.
Restart the unit. If the
error occurs again, notify
Erbe Service.
EIP-A 160 Current too high when
switched ON.
Restart the unit. If the
error occurs again, notify
Erbe Service.
EIP-A 192 Motor control defective
(current too high when
switched ON).
Restart the unit. If the
error occurs again, notify
Erbe Service.