6 • Remedying malfunctions
30 / 84
80116-938_V24392
2023-03
G/G-A error
Display on the unit: G or G-A, detecting module: UGM
Error code Description Additional information Action
G 1 CRC check triggers an
alarm.
Restart the unit. If the
error occurs again, notify
Erbe Service.
G 5 Oscillator clock fault. From V 1.1.x. Restart the unit. If the
error occurs again, notify
Erbe Service.
G 6 Oscillator fault. From V 1.1.x. Restart the unit. If the
error occurs again, notify
Erbe Service.
G 10 D-Flash programming
error.
Restart the unit. If the
error occurs again, notify
Erbe Service.
G 11 P-Flash programming
error.
Restart the unit. If the
error occurs again, notify
Erbe Service.
G 75 Temperature error. Generator: Temperature
too high!
Allow the unit to cool
down. Activation may only
be repeated once the unit
has cooled down.
If the error occurs again,
notify Erbe Service.
G-A 75 Temperature error. Generator: Temperature
too high!
Allow the unit to cool
down. Activation may only
be repeated once the unit
has cooled down.
If the error occurs again,
notify Erbe Service.