W-/W-A error
Display on the unit: W or W-A, detecting module: GWM
Error code Description Additional information Action
W 1 – 8 Checksum partition x. Restart the unit. If the
error occurs again, notify
Erbe Service.
W 9 – 16 No memory partition x. Restart the unit. If the
error occurs again, notify
Erbe Service.
W 17 – 24 Write/read error
partition
x.
Restart the unit. If the
error occurs again, notify
Erbe Service.
W 28 – 30 Restart the unit. If the
error occurs again, notify
Erbe Service.
W 32 Update error. Restart the unit. If the
error occurs again, notify
Erbe Service.
W 33 Update procedure in
progress.
Wait until the GWM
update is complete.
W 36 No more RAM available. Software. Restart the unit. If the
error occurs again, notify
Erbe Service.
W 37 RAM error. Hardware. Restart the unit. If the
error occurs again, notify
Erbe Service.
W 40 CPU module overheating. Hardware. Switch off unit and allow
to cool down. If the error
occurs again => notify
Erbe Service.
W 41 Faulty CPU module volt-
age regulator.
Hardware. Restart the unit. If the
error occurs again, notify
Erbe Service.
W 44 Container error. Software. Restart the unit. If the
error occurs again, notify
Erbe Service.
W 48 + 49 Restart the unit. If the
error occurs again, notify
Erbe Service.
W 52 + 53 Restart the unit. If the
error occurs again, notify
Erbe Service.