6 • Remedying malfunctions
32 / 84
80116-938_V24392
2023-03
I-A 34 Activation ID – malfunc-
tion.
Check the accessories. If
the error persists, notify
Erbe Service.
I-A 35 SK error, undefined:
In the case of a monopo-
lar or bipolar socket, or of
an MF-U socket, the left
and/or center pin is
inserted but the one on
the right is not.
Check the accessories. If
the error persists, notify
Erbe Service.
I-A 36 SK error, bipolar:
2 pins detected for each
plugging detection at
monopolar socket (cor
-
rect: 1 or 3).
Check the accessories. If
the error persists, notify
Erbe Service.
I-A 37 SK error, monopolar:
3 pins detected at bipolar
socket
(correct: 2 – right + 1
additional).
Check the accessories. If
the error persists, notify
Erbe Service.
I-A 38 Protocol version not com-
patible with 1-wire.
From V 1.1.x. Check the accessories. If
the error persists, notify
Erbe Service.
I-A 40 One or more buttons
detected as "pressed" on
insertion.
Check the accessories. If
the error persists, notify
Erbe Service.
Display on the unit: I or I-A, detecting module: IIM
Error code Description Additional information Action