6 • Remedying malfunctions
36 / 84
80116-938_V24392
2023-03
M-A 1 thermoSEAL short circuit:
No tissue effect, release
the hand trigger of the
instrument if required.
Can only occur in the case
of BiCision due to an elec
-
tromechanical fault.
Check the accessories.
M-A 2 thermoSEAL (no load):
Insufficient tissue contact.
Grip the tissue again.
Make sure that there is
sufficient tissue contact.
Check the accessories.
M-A 10 Problem when perform-
ing the initial incision
using highCUT bipolar:
Initial incision could not
be completed success
-
fully!
Repeat activation and
quickly guide the loop
towards the tissue.
Check the accessories.
M-A 11 highCUT, bipolar (no load):
Instrument not dipped
into NaCl solution or
check contact.
Ensure that NaCl solution
is used as the irrigation
solution.
Check the accessories.
M-A 20 Power overdose in the
case of PPS/QuickStart.
Guide the instrument
quickly to the tissue. If
possible, switch off the
QuickStart function (via
the effect window of the
mode being used).
M-A 30 endoCUT Q idle state:
No tissue contact was
detected for the duration
of two initial incision
cycles.
Place the instrument in
contact with the tissue
and reactivate.
If the error persists, check
the footswitch assign
-
ment. Check instrument
and connecting cable..
Display on the unit: M or M-A, detecting module: MCM
Error code Description Additional information Action