EasyManua.ls Logo

Erbe VIO 3 - Page 39

Erbe VIO 3
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39 / 84
6 Remedying malfunctions
80116-938_V24392
2023-03
N-A 55 NESSY: Split return elec-
trode connected.
Select the type of return
electrode used. If the
message persists, check
the return electrode
accessories.
N-A 190 NESSY: Symmetry warn-
ing.
Check the alignment of
the return electrode. Fol
-
low the information pro-
vided in the User Manual.
If the error persists, check
the return electrode
accessories.
N-A 191 NESSY: Current density
warning.
A higher temperature is
possible under the return
electrode! Check whether
the activation times and
the effect setting can be
reduced.
N-A 192 NESSY: HF current warn-
ing. NESSY current of
300
mA exceeded.
NESSY: HF current warn-
ing (neonatal).
A higher temperature is
possible under the return
electrode! Check whether
the activation times and
the effect setting can be
reduced.
Display on the unit: N or N-A, detecting module: NM
Error code Description Additional information Action

Table of Contents

Other manuals for Erbe VIO 3

Related product manuals