6 • Remedying malfunctions
44 / 84
80116-938_V24392
2023-03
S-A 30 Two identical
footswitches have been
detected.
Disconnect both
footswitches from the
unit. It is only possible to
connect one two pedal
footswitch and one one
pedal footswitch in each
case.
S-A 31 A modified module con-
figuration was detected.
-> internal modules /
IIM X.
Check whether all config-
ured modules in the sys-
tem have been correctly
detected.
If the error occurs again,
notify Erbe Service.
S-A 32 Occurs when a modified
socket configuration has
been detected. (e.g. due
to replacement or retrofit
-
ting of a socket or in the
case of a socket type
detection defect)
Check whether all config-
ured sockets in the sys-
tem have been correctly
detected. If correct, con
-
firm the dialog shown.
If the error occurs again,
notify Erbe Service.
S-A 40 Two pedal FS footswitch
pedal (CUT, COAG or
ReMode) was reported as
pressed during the first
query.
Check the two pedal
footswitch. If the error
persists, notify Erbe Ser
-
vice.
S-A 41 One pedal FS footswitch
pedal (COAG or ReMode)
was reported as pressed
during the first query.
Check the one pedal
footswitch. If the error
persists, notify Erbe Ser
-
vice.
S-A 129 Heat management: High
average power output
reached.
The unit has heated up
considerably. Minimize
the activation time and, if
possible, reduce the effect
setting.
S-A 130 The instrument that is
plugged in is blocked or it
is not supported and can
-
not be used.
Check the accessories.
Disconnect incompatible
instrument from the unit.
S-A 149 Double footswitch activa-
tion.
Check the two pedal
footswitch. If the error
persists, notify Erbe Ser
-
vice.
Display on the unit: S or S-A, detecting module: SCM
Error code Description Additional information Action