S2000 Elite - User's Guide
035-00707 Revision 1
88
Output
Virtual Contact (VO)
1 to 32
This virtual output is
an internal contact
that is synchronized
with the
corresponding virtual
input. For instance, if
an instruction is to
energize Virtual
Output #1, Virtual
Input #1 contact will
be closed. If an
instruction de-
energizes Virtual
Output #1, Virtual
Input #2 contact will
open
See Virtual Inputs
and Outputs for more
details on how to use
the virtual
inputs/outputs
This output can be
latched
(energized)/unlatched
(de-energized)
Output
Exposure Action (%, W, W/cm
2
)
0 to 100%
if intensity
type, or
value is
limited by
the
calibration
The exposure is a
special action where
it will run an exposure
when the action
executes
The specific
StepCure instruction
with a running
exposure will be
momentarily blocked
until that exposure
completes
The other sequenced
steps will still execute
Only one exposure
action can execute at
a time