S2000 Elite - User's Guide
035-00707 Revision 1
87
(energized)/unlatched
(de-energized)
See PLC Connector:
DB50 for more details
on the pin out
Output
Tree 1/2/3/4 (T1/T2/T3/T4)
N/A
This output is a
physical output that
can be used by the
process line to
control external
equipment
This output can be
latched
(energized)/unlatched
(de-energized)
See PLC Connector:
DB50 for more details
on the pin out
Output
Duration (DU)
0.1 to
6553.5s
The duration is a
special output where
the action performed
is to load a timer with
the specified value in
seconds
The duration block
must be followed by a
virtual output contact
or an exposure
action, for the timer to
become activated
If the duration block
is before an exposure
action, the duration
value will set the
exposure duration
If the duration block
is before a virtual
output, the duration
value will set the time
delay until the virtual
output is activated