Operator‘s Manual FERITSCOPE
®
FMP30 Page 27
System Setup, Maintenance and Cleaning
Fig. 3-4 Connecting a probe
5.
Use ON/OFF to turn the instrument on again. The instrument automati-
cally recognizes the type of probe connected to it.
Exception: A flashing symbol for the measurement method on the display
indicates that the instrument does not recognize the connected probe. In
such a case:
A new Application must be set up for the connected probe
( 5.1 ‘Setting Up an Application’, beginning on Page 37) and/or
The probe must be assigned to the instrument or to the respective Appli-
cation, respectively ( 4.2 ‘Assigning a New Probe’, beginning on
Page 35) or
The probe that has been linked to the respective Application thus far
must be reconnected.
.
A new corrective calibration must be performed after the probe has been
assigned (
Page 64)!