EasyManua.ls Logo

Fisher FERITSCOPE FMP30 - Malfunctions; Malfunctions and Messages

Fisher FERITSCOPE FMP30
241 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator‘s Manual FERITSCOPE
®
FMP30 Page 187
Malfunctions and Messages
11 Malfunctions and Messages
11.1 Malfunctions
Malfunction/Message Cause Correction
start
on
Page
No display Instrument not switched
on.
Use ON/OFF to turn the
instrument on.
28
Instrument switched off
automatically (energy
savings function during
operation without AC
adapter
Use ON/OFF to turn the
instrument on.
28
Batteries are dis-
charged (operation with-
out AC adapter)
Replace rechargeable
or regular batteries with
new ones or operate
the instrument with the
AC adapter.
23
No display of the re-
sults with
FINAL-RES
No readings in the cur-
rent Application
(e.g., because the mea-
surement data have
been deleted).
Take measurements. 85
No display of the re-
sults with BLOCK-
RES
No readings in the
current Application
(e.g., because the mea-
surement data have
been deleted).
Take measurements. 85
No change on the
display for
ZERO, CAL or
MENU
Restricted operating
mode is enabled.
Use ON/OFF + EN-
TER to disable the re-
stricted operating mode
with the instrument
switched off before-
hand.
166