Malfunctions and Messages
11.2 Messages on the Display
The error messages (E***) and warning messages (W***) that may occur
during instrument operation are contained in the overview on the following
pages.
Malfunction/Message Cause Correction
start
on
Page
Reading cannot be dis-
played (because the val-
ue is greater than 9999
or smaller than
- 9999)
Cause: Measurement
carried out incorrectly.
Perform the measure-
ment correctly (e.g., do
not allow the probe to
hover over the speci-
men before or after the
measurement; observe
the minimum lift-off dis-
tance!)
34
Cause: Eddy current
probe that is currently
normalized on a non-fer-
romagnetic base materi-
al has previously been
normalized on a ferro-
magnetic base material.
Continue normaliza-
tion.
61
Instrument is not ready
to make measurements.
Cause: Re-initialization
was carried out.
Application was delet-
ed and then switched
off.
Use ZERO to start and
perform the normaliza-
tion.
61
Histogram does not print
because fewer than 30
readings are stored in
the current Application.
Take more than 30
measurements.
85
- - - -
E 000
Not enough
measurements
!