Page 184 Operator‘s Manual FERITSCOPE
®
FMP30
Instrument Settings - Service Function Menu
10.8 Storage Mode
10.9 Performing a Master Calibration
Performing the master calibration: 6.4 ‘Master Calibration’, beginning
on Page 70.
Key sequ. /
Action
Detail of the display Explanation
Select Storage mode by pressing the ar-
row key
or and confirm the selection
with ENTER.
ENTER Select the desired mode by pressing the ar-
row keys
or and confirm the selection
with ENTER.
[store]: The readings are stored and will
be retained even after the instrument is
switched off
[do not store]: The readings will be dis-
played but not stored
[delete at off]: All readings will be deleted
when the instrument is switched off
Use DEL at any time to cancel the setting
procedure.
Key sequ. /
Action
Detail of the display Explanation
Select Master calibration by pressing the
arrow keys
or and confirm the selec-
tion with ENTER.