Page 36 Operator‘s Manual FERITSCOPE
®
FMP30
Probe Handling
Key sequ. /
Action
Detail of the display Explanation
This warning appears briefly after power-up
if a probe other than the last one used is
connected to the instrument. After that, the
display of the measurement method flashes.
4.
ZERO
Example: FGAB1.3F = Name of the con-
nected probe
DEL: Probe assignment starts
ENTER: Probe will not be assigned, mea-
surement method display flashes
5.
DEL
DEL: Probe will be assigned to all Applica-
tions
ENTER: The probe will be assigned only to
the current Application
6.
DEL
DEL: All stored readings will be deleted;
the probe will be assigned to the current
Application / all Applications
ENTER: The probe will not be assigned,
measurement method display continues to
flash
7.
ZERO appears on the display.
A normalization is required
( beginning
on Page 61).
[Base material (Fe)]: The measurements
for the normalization should be made on
base.
ENTER: Cancels the normalization.