Operator‘s Manual FERITSCOPE
®
FMP30 Page 59
Applications
Immediately following the linking procedure:
Applications 4 and 6 are also linked to each other , because
they have been set up with the same FGABW1.3-FE probe (serial num-
ber 505).
Even though Applications 1 and 3 have been set up with the same
FGAB1.3-FE probe (serial number 101), they are not linked to each
other because the readings in Application 1 are displayed in ferrite
percent [%] and the readings in Application 3 in ferrite numbers [FN].
Application 5 is not linked to these two Applications because it has
been set up with the FGAB1.3-FE probe with the serial number 909.
Application 2 is not linked to any other Application because no other
Application has been set up with the FGABI1.3_150FE probe (serial
number 303).
After the corrective calibration of Application 6:
Because Applications 4 and 6 are linked with each other, after the
corrective calibration of Application 6, the new corrective calibration
(Norm 6n, Cal 6n, grayed out in Fig. 5-3) will be stored in Applica-
tion 4 as well. The previous normalization and corrective calibration
(Norm 4/6 and Cal 4/6) will be overwritten.
Because a normalization or a corrective calibration has not yet been
carried out immediately following the linking procedure, the normal-
izations and corrective calibrations of the linked Applications still dif-
fer from each other.
The normalization and corrective calibration of all other Applications
remains unchanged in spite of linking, because a normalization or
corrective calibration has not yet been carried out with these probes.