5502A
Operators Manual
6-8
Table 6-8. Setup and Utility Commands
Command Description
CLOCK Sets the real-time clock.
CLOCK? Queries the real-time clock.
DBMZ_D Sets the power-up and reset default impedance used for dBm outputs (ac volts).
DBMZ_D? Shows the power-up and reset default impedance used for dBm outputs (ac volts).
FORMAT
Use with extreme care. Restores the contents of the nonvolatile memory device to
factory defaults.
LIMIT Sets the maximum permitted output magnitudes, negative and positive.
LIMIT? Shows the programmed output magnitude limits for voltage and current.
PR_RPT
Prints the Stored, Active or CAL-Constant CAL_Report through the HOST or UUT Serial
Port.
RTD_TYPE_D Set the default Resistance Temperature Detector (RTD) sensor type.
RTD_TYPE_D? Shows the default Resistance Temperature Detector (RTD) sensor type.
SP_SET
Sets the HOST serial port communication parameters and saves them in nonvolatile
memory.
SP_SET?
Shows the HOST serial port communication parameters contained in nonvolatile
memory.
TC_TYPE_D Sets the power-up and reset default thermocouple type.
TC_TYPE_D? Shows the power-up and reset default thermocouple type.
TEMP_STD Sets the temperature degree standard, IPTS -68 or its-90.
TEMP_STD? Shows the temperature degree standard, IPTS -68 or its-90.
TLIMIT_D
Sets the power-up and reset default for the time limit for –SC600 OVERLD mode to stay
in operate.
TLIMIT_D?
Shows the power-up and reset default for the time limit for –SC600 OVERLD mode to
stay in operate.
UNCERT?
Shows specified uncertainties for the current output. If there are no specifications for an
output, shows zero.