Remote Commands
Commands
6
6-31
A reset gets the subsequent commands and values:
Command Value Command Value
CUR_POST AUX RTD_TYPE <RTD_TYPE_D value>
DBMZ <DBMZ_D value> SCOPE OFF
DC_OFFSET 0V STBY (No output)
DUTY 50PCT TC_OFFSET 0 CEL
EARTH OPEN TC_OTCD ON
HARMONIC 1, PRI TC_REF INT
LCOMP OFF TC_TYPE <TC_TYPE_D value>
LOWS TIED TRIG OFF
OUT 0V,0HZ
Command Value Command Value
OUT_IMP Z1M WAVE NONE,NONE
PHASE 0DEG ZCOMP OFF
RANGELCK OFF ZERO_MEAS OFF
TSENS_TYPE TC
Changes made to the setup menus that are not kept in memory are discarded on reset.
Response: (None)
Example: *RST
Put the Calibrator in a reset condition, evoking the commands and values shown above.
RTD_TYPE
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
(Resistance Temperature Detector Type command) Sets the Resistance Temperature
Detector (RTD) sensor type.
Before you use RTD_TYPE, select RTD with the TSENS_TYPE command. After you use
RTD_TYPE, select the output temperature with the OUT command. Changes in
temperature sensors changes the output to 0 °C. When set, the Calibrator keeps the RTD
type until power off or reset.
Parameters: PT385 (100-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_200 (200-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_500 (500-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_1000 (1000-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT3926 (100-ohm RTD, curve α=0.003926 ohms/ohm/°C)
PT3916 (100-ohm RTD, curve α=0.003916 ohms/ohm/°C)
CU10 (10-ohm RTD, empirical curve)
NI120 (120-ohm RTD, empirical curve)
Example: RTD_TYPE PT3926
Set the RTD type to a 100-ohm type, with the pt3926 curve (α=0.003926 ohms/ohm/°C).
The resistance of 100 ohms refers to the ice point property, (the resistance of the RTD at
0 °C (32 °F)).